Operator’s Manual · Version 3.0 145
Multiplate® analyzer 5 Operation
Shutting down the analyzer
Shutting down the analyzer
Permanent use The analyzer is designed for permanent use 24 hours a day and can be left switched
on. It is, however, recommended to shut it down and switch it off once a week.
Perform an electronic control measurement daily and after a weekly shutdown.
NOTICE
Loss of data due to inappropriate shutdown
Shutting down the system or switching off the analyzer while tests are being performed
may lead to data loss.
r Do not shut down the system or switch off the analyzer while processing samples.
r Stop all measurements and shut down the system.
r Do not use the power button to shut down the system.
p To shut down the analyzer
1
Choose Multiplate > Exit. (You can also click in the top right corner of the
main screen.)
A confirmation window is displayed.
2
In the confirmation window, choose the Yes button.
The software is shut down. When the shutdown is complete, the Windows screen
is displayed.
3
In the Windows task bar, choose Start > Shut down.
A confirmation window is displayed.
4
Confirm the shutdown.
The Windows operating system is shut down and the analyzer is switched off.
s
Taking the analyzer out of operation
No special action is required, when the analyzer has to be taken out of operation
when it is not used for a longer period of time. Shut down the analyzer and unplug
the power plug.