Multiplate® analyzer 5 Operation
Performing a measurement
15
Disconnect the sensor cables from the test cells.
16
Remove the test cells and discard them according to the local regulatory
standards.
s
Sending results to the LIS
If automatic result upload is disabled, the results in the worklist have to be manually
sent to the LIS after validation.
Discarding orders Use the Discard button to reject an order that cannot be measured for any reason.
Discarded orders are sent back to the LIS.
q
To delete an oder use the Delete button. In this case, the order is removed from the
worklist without sending it back to the LIS.
p To send results to the LIS
1
Validate the results.
2
Select the orders that you want to send to the LIS.
Only orders with final results can be sent to the LIS.
3
Choose the Send button.
The results of the selected orders are sent to the LIS.
s
p To discard orders
1
Select the orders that you want to discard.
Only orders that are not assigned to a channel can be discarded.
2
Choose the Discard button.
Discarded orders are sent with an empty result to the LIS and removed from the
worklist.
s
Performing measurements using the electronic pipette
The electronic pipette simplifies and standardizes the measurement process,
particularly sample handling, sample assignment, and pipetting. The operator is
guided through the procedure by several “next pipette step” instructions.
q
o The electronic pipetting can be canceled any time by clicking in the top right
corner of the Process measurement series window. When canceling, all pipettings
and used test cells are lost. When measurements are not yet started, canceling the
electronic pipetting also will delete all input data.
o Roche recommends to use the electronic pipette. Use manual pipetting only in case of
problems with the electronic pipette.