Multiplate® analyzer 6 Configuration
Configuring the analyzer
q
10 AU*min = 1 U
General analyzer configurations
f Configurati on > Edit analyzer configuration > Configurations
The options on the Configurations tab can be changed with a level 2 password only.
Figure 6-9 Configurations tab
Max. measurement duration The default duration for the standard test measurement is 6 minutes. This period
may be adjusted to between 3 to 30 minutes. If you enter an invalid value (< 3 min or
> 30 min) a warning is displayed, informing you of this fact.
Lab ID A laboratory ID can be added to all printouts by entering details into this window.
Timer runtime The timer runtime default is 180 seconds, it can be adjusted between 1 and 900
seconds.
Specified temperature The specified temperature of the analyzer is 37.0 °C. The tolerance range is
37.0±1.0 °C. The value may be changed for research purposes.
Min. specified temperature Default value is 36.0 °C. The value may be changed for research purposes between
20.0 and 45 °C.
Max. specified temperature Default value is 38.0 °C. The value may be changed for research purposes between
20.0 and 45.0 °C.
Tur n off heating above Do not change this value!
If the device reaches a temperature of more than 42.0 °C the software turns off the
heating to prevent damage to electronic parts.
Waiting time in minutes after the
start of the software until an
electronic QC can be started
Do not change this value!
The waiting time of 20 minutes ensures that the measurement block reaches the
specified temperature before an electronic control measurement can be started.