Operator’s Manual · Version 3.0 177
Multiplate® analyzer 7 Maintenance
Maintenance and cleaning of the electronic pipette
14
Insert the spring and seal.
Figure 7-17 Inserting spring with seal
15
Screw on the retention sleeve. (Hand-tight is sufficient).
Figure 7-18 Screwing on retention sleeve