Roche Diagnostics
Operator’s Manual · Version 3.0 179
Multiplate® analyzer 7 Maintenance
Maintenance and cleaning of the electronic pipette
18
Screw the pipette shaft into the grip. (Hand-tight is sufficient.)
Figure 7-21 Screwing pipette shaft on grip
19
Insert a new tip cone filter into the pipette.
Figure 7-22 Inserting a new filter
20
Perform a reference run.
u For details, see To perform a reference run (p. 180).
When the reference run is finished, the display automatically returns to the
normal view.
Figure 7-23 Normal display view