194 Operator’s Manual · Version 3.0
8 Troubleshooting Multiplate® analyzer
Dealing with exceptional situations
Stirring module messages
Situation Description Cause Remedy
Message window
shows The stir bar
monitoring device
does not respond.
Do you want to
continue?
The stirring control
sends an error or is
not working.
The stirring control might be defective. 1. Choose the No button to stop the
measurement.
2. Repeat the measurement.
3. If the error persists, contact your
Roche Service representative.
Message window
shows The magnetic
induction of the
stirring mo dule lies
outside the specified
range.
Do you want to
continue?
The magnetic
induction signal of
the stirring module is
outside of the default
range.
The stirring module might be defective. 1. Choose the No button to stop the
measurement.
2. Repeat the measurement.
3. If the error persists, contact your
Roche Service representative.
Message window
shows The speed of
the stir bar li es
outside the specified
range.
Do you want to
continue?
The stirring speed of
the stirring module is
outside of the default
range.
The stirring module might be defective. 1. Choose the No button to stop the
measurement.
2. Repeat the measurement.
3. If the error persists, contact your
Roche Service representative.
Table 8-5 Stirring module messages