100
77 ApplicationsApplications
5. Set the Solvent
(1) Select [Solvent], and then press [ OK].
(2) Select either [Water], [Ethyl alcohol] or [Arbitrary] as
the liquid medium, and then press [ OK].
(3) Set the [Solvent temperature], and then press
[ OK].
(4) Enter the temperature of the solvent between a
range of 10.0 to 30.0°C, and then press [ OK].
(5) Press [ LEFT].
Selecting [User-specified] as the Solvent
If [User-specified] has been Selected in Step (2), select
[solvent Density], press [ OK], enter the density
of the liquid medium within a range of 0.0001 and
999.9999, and then press [ OK].
6.
Set the number of digits to be displayed after the decimal point
for the solid specicity gravity value in accordance
(1) Select [Number of Digits to be Displayed after the
Decimal Point], and then press [ OK].
(2) Enter the number of digits to be displayed after the
decimal point within a range of 1 to 5, and then
press [ OK].
7. Set display hold at ON or OFF in accordance
(1) Select [Display hold setting], and then press
[ OK].
(2) Select [on] or [off], and then press [ OK].
The solid specific gravity value will be held once it has
been determined if the display hold setting has been
set at [ON], and it will continued to be displayed until
[ON] is pressed.
*
*

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu AP W-AD Series and is the answer not in the manual?

Shimadzu AP W-AD Series Specifications

General IconGeneral
BrandShimadzu
ModelAP W-AD Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals