Outputting Data Saved on the Built-in Memory to USB Memories
Timing for Outputting Data to the Built-in Memory
Data is output to the built-in memory on a constant basis, but to the USB memory only when
the setting has specified.
1. Call out the Save memory Settings
(1) Press [ MENU] while in the weight measurement mode, and then press [ LEFT].
(2) Select [
System settings], and then press [ OK].
(3) Select [Memory save setting], and then press [ OK].
2. Set built-in memory output
(1) Select [Internal memory output], and then press
[ OK].
(2) Select any of the output methods, and then press
[ OK]
When the data saved on the internal memory is to be deleted
Select the [Memory Saving Setup] menu, select the
[Delete Internal Memory (Password) (Deletes All Data)], and
then press [ OK]. A prompt requiring the administrator
authorization (password) will be displayed. Enter the
administrator’s password and then press [ OK] to
delete all of the data stored on the internal memory.
^ [Deleting the Contents of the Built-in Memory] (P.87)
Output Method Output Details Output Destination
Setting Data → USB Memory Setting Data USB Memory
Weighing value → USB Memory (All) All weighing values USB Memory
Weighing Value → USB Memory
(Deviations Only)
Deviating weighing values USB Memory
Weighing Value → Serial Communication (All) All weighing values Serial Communications
Weighing Value → Serial Communication
(Deviations Only)
Deviating weighing values Serial Communications
Record of Calibration → USB Memory (All) All records of calibration USB Memory
Record of Calibration → USB Memory
(Deviations Only)
Deviating records of calibration USB Memory
Record of Calibration → Serial Communication
(All)
All records of calibration Serial Communications
Record of Calibration → Serial Communication
(Deviations Only)
Deviating records of calibration Serial Communications
What does [Deviation] mean?
The deviation functions enables the previous data to be recorded so that only the new
measurement data is output.
3. Return to the weight measurement mode
Press [ POWER].
File Names Output to the USB Memory
Files are output with the names [XXXYYYYMMDDhhmmss(_?).file extension].
XXX: WEI = Weighing value, CAL = Record of Calibration, SET = Setting information,
CAP = Weight Screen
YYYY: Year
MM: Month
DD: Date
File extension: txt or CSV
*
*
*