Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
245
Flowrate decimal (2.13.4.6.)
Sets the maximum number of decimal places to be displayed in the auxiliary reading.
No digits
Decimal position (2.12.4.) (Page 237)
The maximum number of decimal places shown in the Primary Reading is controlled by
Decimal Position (2.12.4.) (Page 237).
Flowrate units (2.13.4.7.)
Defines the units used for flow calculations.
Note
Set this parameter only when using BS
-3680/ISO 4359 Rectangular Flume or BS-3680/ISO
-Notch Weir [Primary Measuring Device (2.13.2.) (Page 240) =
-36806 or Thin Plate V-Notch Weir BS-36807]. Use the default option
Ratiometric calculation for Flowrate Units (2.13.4.7.) when Primary Measuring Device
(Page 240) = Exponential devices, Palmer-Bowlus Flume, H-Flume, Universal
Linear Flow Calculation, or Universal Curved Flow Calculation.
[Primary Measuring Device (2.13.2.)
*Ratiometric calculation
[Primary Measuring Device (2.13.2.)
(Page 240) = Rectangular Flume BS-
3680 or Thin Plate VNotch Weir BS-
3680 only]
3
3
FT
3
/S (Cubic feet/second)
IMPGAL/MIN (Gallons/minute – Imperi-
MIMPGAL/D (Million gallons/day –
GAL/MIN (Gallons/minute – U.S.)
MUSGAL/D (Million gallons/day – U.S.)
Flow time units (2.13.4.4.) (Page 244)
Primary measuring device (2.13.2.) (Page 240)
Flowrate decimal (2.13.4.6.) (Page 245)