(De)activates the MultiFlow temperature
measurement.
IMPORTANT: With an inactive sensor no
compensation and no mass
computation can be carried out!
IMPORTANT: If no sensor is connected, the
temperature inputs must be fitted with
a jumper.
9.3.3.2.1
Temperature Sensor (PT 100)
A temperature offset must be defined to compensate for component
tolerances. This temperature offset is determined during the preliminary
test with a number of reference temperatures. The values found are
included in the preliminary certificate and must not be changed.
If during a recalibration it is found that the temperature measured by the
MultiFlow is no longer within the calibration limits, then parameter 3.3.2.1
must be changed appropriately.
9.3.3.2.2
Working Without a Temperature Sensor
If the temperature sensor is disabled (only applicable from version 2.0
and higher) no compensated volumes or weight information (calculation
is always based on compensated volumes!) may be displayed anymore.
These values will not appear in reports (on paper or on the chip card), or
they will be set to "0" or "–". The same applies to the corresponding
totalizers.
If not temperature sensor is used (e.g. for motor spirits for which no
temperature compensation is required) you must take care that bridges
are set at the terminals instead of the temperature sensor (PT100).
See drawings 51.350956 (page 321) and 51.350958 (page 330)