50
Sening® is a registered trademark of FMC Technologies
MN F09 002 US  DOK-383E  Issue/Rev. 3.65 (02/24)
MultiFlow  Initial Operation
5.2.1.1
MultiFlow up to version 5.00
The "Program update" menu item (menu 4.6) has been added to the
"Service" section of the main menu to enable the program transfer via the
printer line in the case of multiple measuring systems. This menu offers
you the functions "Activated" (menu 4.6.1) and "Disabled" (menu 4.6.2).
In the case of vehicles with several measuring systems these functions
enable you to select the specific MultiFlow onto which a new program
should be loaded.
Only the specific MultiFlow on which a program update is to be carried
out may be set to “Activated”; all others must be “Blocked”.
The status currently active is shown in the display (in double-size
lettering). The update mode is terminated by pressing any key.
The manual supplied with the service program contains detailed
instructions for reading in a new program version via the printer line. See
chapter 15.2 "Software update by download" / page 275.
5.2.1.2
MultiFlow from version 5.00
The functions described below for the approval and management of
program updates are available only in the MID MultiFlow variant (from
version 5.00).
Program update menu
All software download attempts
involving changes in the
calibration-relevant
programming section - whether
successful, faulty or called off
are saved in a special logbook
for traceability purposes. The
maximum number of admissible
downloads is limited to 30,
wherein only downloads with
changes in the calibration-
relevant program section are
logged. Downloading a software
in which only changes in the non-calibration relevant program section
have been carried out does not result in a reduction in the still available
download attempts. No further downloads can be carried out should the
maximum number be attained or the logbook damaged. In such an
Pro gr am u p da te
S el ec ti on : 46
1 Pr og ra m u pd at e
2 Re ma in . at t emp ts
3 U pd at e re por t
4 R es et
For wa rd B ac k - --