Configure panel (Measurement configuration menu)
Use the Configure panel to set the measurement source(s), add a custom name (label) for the measurement, and other parameters.
To open the Configure panel:
1. Double-tap a Measurement badge to open the Measurement configuration menu.
2. Tap the Configure panel.
Not all items listed are shown for all measurements; The panel only shows fields and controls relevant to the selected measurement.
Field or control Description
Source Sets the signal source used to take the measurement. Tap the field to show the list of available
sources. If the measurement requires more than one source, multiple Source fields are displayed.
Label Sets the name of the measurement. You can use the default name, or double-tap in the field and
change the label using a connected keyboard or the virtual keyboard.
Signal Type Sets the signal type (Clock, Data, Auto) of the source signal for some measurements.
Source 1 Sets the Source 1 for the measurement where two sources are required. Available for Delay, Phase,
and Skew measurement.
Source 2 Sets the Source 2 for the measurement where two sources are required. Available for Delay, Phase,
and Skew measurement.
Edge, Clock edge Sets the edge of the signal to use for starting the measurement.
From Edge Sets the Source 1 waveform edge on which to start the measurement, for two-source measurements.
To Edge Sets the Source 2 waveform edge on which to stop the measurement, for two-source measurements.
Calculate One Measurement
Per
Sets the amount of waveform data to use to calculate one measurement; one measurement across the
entire waveform record or one measurement for each cycle of the waveform in the record.
Other measurement panels
Measurement Name panel (Measurement configuration menu) on page
115
Reference Levels panel (Measurement configuration menu) on page 116
Gating panel (Measurement configuration menu) on page 118
Pass/Fail Testing panel (Measurement Settings menu) on page 119
Reference Levels panel (Measurement configuration menu)
Use the Reference Levels panel to set the scope of the reference level settings (global or local), the reference levels (High, Mid, and Low),
the units used to take measurements, and the method used to calculate the Top and Base waveform values. You can set the levels to be
the same or different for rising and falling edges.
To open the Reference Levels panel:
1. Double-tap a Measurement badge.
2. Tap the Reference Levels panel.