5. T
ap the Time Limit field to set the time value for the Trigger When condition.
6. Tap the Upper Threshold field and use the assigned multipurpose knob (or double-tap the field and use the keypad to enter a value)
to set the length of setup time that the data signal must be stable before a clock edge occurs.
7. Tap the Lower Threshold field and use the assigned multipurpose knob (or double-tap the field and use the keypad to enter a value)
to set the length of hold time that the data signal must be stable after a clock edge occurs.
8. Use the Slope controls to set the signal transition direction to detect. (rising, falling, or either direction).
Set up trigger on a parallel bus
Use this procedure to set up triggering on a parallel bus.
Use this procedure if you have already created a parallel bus.
1. Double-tap the Trigger badge.
2. Tap the Trigger Type field and select Bus from the list.
3. Tap the Source field and select the parallel bus on which to trigger.
Advanced triggering