Calibration configuration menu fields and controls
Field or control Description
Factory Adjustment Status This area at the top of the menu lists the instrument calibration status. Factory Adjustment Status
should be Calibrated.
If an instrument becomes uncalibrated, this displays an Uncalibrated status and a red W
arning
message bar is shown at the top of the screen in the Menu bar area. Contact your nearest Tektronix
Service Center for assistance.
SPC Status Indicates the status of the last SPC run (Pass or Failed). Also indicates how long ago the last SPC was
run.
Run SPC Signal path Compensation (SPC) corrects for internal DC inaccuracies caused by temperature
variations and/or long-term drift in circuits.
Note: SPC takes three minutes per channel to run.
Allow the instrument to warm up for 20 minutes before running SPC.
Remove all probes, cables, and adapters from the input connectors before running SPC.
T
ap Run SPC to run signal path compensation. The compensation can take up to three minutes per
channel.
Security configuration menu
Use this menu to enable or disable LAN port and securely erase the oscilloscope nonvolatile memory (T
ekSecure®). Access to the
functions on this menu are password protected.
To open the Security configuration menu:
1. Tap Utility in the Menu bar.
2. Tap Security.
3. Tap On or Off to enable or disable LAN port functions. The default setting is On.
4. Tap Set Password to enter a password for the first time. If a password has already been set, tap Change Password to change the
password.
Note: The switches cannot be toggled (of
f or on) unless a password has been set.
• If a switch is toggled without a password set, a message is shown asking you to set a password.
•
If a switch is toggled with a password set, the instrument shows the Enter Password menu.
Security configuration menu fields and controls
Field or control Description
LAN Port Tap to enable (On) or disable (Off) the Ethernet port.
Set Password Opens the Set Password menu.
Change Password Opens the Change Password menu.
Table continued…
Menus and dialog boxes
2 Series MSO MSO24 and MSO22 197