Probe Setup panel (Channel configuration menu)
Use the Channel configuration menu Probe Setup panel to setup probe, deskew
.
To open the Probe Setup panel:
1. Double-tap an analog Channel badge on the Settings bar to open the Channel configuration menu.
2. Tap the Probe Setup panel.
Probe Setup panel fields and controls
Available fields and controls vary with the type of probe that is attached. For more information, consult the probe documentation.
Field or control Description
Probe Type Sets the probe type as V
oltage or Current.
Deskew Sets or displays the probe deskew value.
Set to 0 Sets the probe deskew value to zero (0) seconds.
Attenuation Set attenuation as 1X, 10X, or Custom.
Custom Attenuation Set the value for Custom Attenuation. Available when Attenuation is Custom.
Measure Current Measures the current. Available when Probe Type is Voltage.
Ratio Double-tap the numeric fields to set the ratio using the virtual keypad. Available when Probe Type is
Voltage and Measure Current is Yes.
Set to Unity Sets the ratio to unity. Available when Probe Type is Voltage and Measure Current is Yes.
Probe error messages are also displayed in this setup panel, see Error Messages and Badges.
Invert channel guidelines
This topic describes how the channel Invert function interacts with other instrument settings.
Settings that change when Invert is set to On or Off:
The instrument clears acquisitions and restarts triggering
All persistence information resets/clears, including all waveform statistics, etc
All associated views and derived results change. For example, XY view, cursor readouts, measurement results, search results, math,
bus decode, and so on, all change to reflect the current Invert setting
Settings that change when Invert is set to On:
The instrument adds a down arrow to the title row of the channel badge to indicate inverted mode
The channel Offset value reverses polarity
The trigger level value reverses polarity
The trigger level indicator moves to the inverted value and the trigger level bar appears momentarily to show the new level
The trigger Slope indicator on the Trigger badge and in its menu changes to the opposite edge (Rising to Falling, or Falling to Rising).
If the Either edge mode is selected then slope mode does not change
Settings that don't change when Invert is On:
The channel Vertical scale does not change
The channel Vertical position knob rotation direction does not change
Menus and dialog boxes
162