3. The second way to delete a Measurement or Search badge is by flicking it of
f the right edge of the display. Flicking to the left from the
right edge of the display recovers the badge.
Note: Badge recovery is possible within 10 seconds only of removal.
Display an XY plot
Use this procedure to display an XY plot.
1. Double-tap Horizontal badge on the Setting bar
.
2. Tap the Acquisition Settings panel.
3. Tap XY to On. The plot is added to the screen.
4. Double-tap in the Plot view to open a configuration menu for that plot.
Display an FFT math waveform
Use this procedure to display an FFT math waveform.
The FFT process mathematically converts the standard time-domain signal (repetitive or single-shot acquisition) into its frequency
components. The FFT function processes the waveform record and displays the FFT frequency domain record, which contains the input
signal frequency components from DC (0 Hz) to ½ the sample rate (also called the Nyquist frequency).
1. Tap Add New Math to create a math waveform and open the Math configuration menu.
2. Tap Source and select the signal source from the list.
3. Set Math Type to FFT. The FFT of the waveform is displayed in an FFT Math waveform view.
4. Double-tap on the FFT waveform display to open a configuration menu to further refine the FFT display.
Add a serial bus to the Waveform view
Use this procedure to add a serial bus to the Waveform view.
This instrument supports decoding parallel buses (standard with the instrument) and several serial buses options. All serial bus functions
are options that must be purchased and installed before they are available in the instrument menus.
Use a Bus configuration menu to define a bus from which to acquire, decode, and display data.
1. Tap the Math Ref Bus > Add New Bus button on the Settings bar to add a Bus badge to the Settings bar, add a bus waveform to the
screen, and open the Bus configuration menu. The default bus type is parallel.
Analog channel operating basics