• Each column header in the file indicates the column of data to be generated from a corresponding pin. If there are no column headers
are present, b0, b1, b2, b3 are added by default. The bits can be in any order as long as column headers are provided.
•
Any comments added must be in English.
• If any of the data bits are missing from the file, it will not load and displays an error message.
• The file size must be less than or equal to 4096 bytes.
Note: The instrument displays an error message if the load file is empty
, corrupted, or the file size is greater than 4096 bytes.
Plot configuration menus
Use plot configuration menus to change settings of a displayed plot. Double-tap a plot view to open its configuration menu.
Plot XY configuration menu
Use this menu to change settings of a displayed XY plot.
T
o open the XY plot menu, double-tap anywhere in the XY Plot view.
Settings panel (Plot XY configuration menu) fields and controls
Field or control Description
X-Axis Sets the source for the X-axis signal.
Y-Axis Sets the source for the Y-axis signal.
• Use the vertical scale and position controls of the waveform connected to the X- and Y-axis sources to set horizontal scale and position
of the XY waveform.
•
XY format is particularly useful for studying phase relationships between two similar signals, creating Lissajous Patterns.
• XY format is a dot-only display, although it can have persistence. The Vector style selection has no effect when you select XY format.
Save panel fields and controls
Use the Save panel controls to save the plot image or date to a file for inclusion in reports or further analysis in other applications.
Field or control Description
Save Plot Image Tap to open the Save As menu. Navigate to the location where you want to save the plot image. Enter
a file name in the File Name field. Select the image file format (PNG, BMP
, or JPG).
Tap OK to save the plot image.
Save Plot Data Tap to open the Save As menu. Navigate to the location where you want to save the plot data. Enter a
file name in the File Name field.
Tap OK to save the plot data as a comma-separated values (.csv) file.
Closing a plot view
T
o close (delete) a Plot view, tap the X in the upper right corner of the view.
Deleting the Measurement badge that enabled the plot also closes the plot.
Menus and dialog boxes
206