Installing your instrument
Operating requirements and setup information to begin using your instrument.
Install option upgrade licenses
Option license upgrades are field-installable licenses that you can purchase after receiving your instrument, to add features to your
oscilloscope. You install option upgrades by installing license files on the oscilloscope. Each option requires a separate license file.
Before you begin
These instructions do not pertain to options that were purchased and pre-installed on your instrument when ordered.
A node-locked license is valid only for the specific model number and serial number of the instrument for which it was purchased; it will not
work on any other instrument. The single license file does not affect options that were factory installed or any other upgrades that you may
have already purchased and installed.
About this task
Note: Y
ou can only install a Node Locked option license one time. If you need to reinstall an uninstalled Node Locked license,
contact Tektronix Customer Support.
Procedure
1. Follow the instructions you received to download the upgrade license file (<filename>.lic).
2. Copy the license file or files to a USB memory device.
3. Insert the USB drive into the powered-on oscilloscope for which the upgrade was purchased.
4. Select Help > About.
5. Select Install License to open the Browse License Files dialog box.
6. Find and select the upgrade license file to install.
7. Select Open. The oscilloscope installs the license and returns to the About screen. V
erify that the installed license was added to the
Installed Options list.
8. Repeat steps 5 through 7 for each upgrade license file that you purchased and downloaded.
9. Power cycle the oscilloscope to enable the installed upgrades.
10. If you installed a bandwidth upgrade, rerun signal path compensation (SPC). Then carefully remove the model/bandwidth label from
the lower-left corner of the front panel and install the new model/bandwidth label that was sent through normal mail channels as part
of the upgrade purchase.
Operating requirements
Use the instrument within the required operating temperature, power, altitude, and signal input voltage ranges to provide the most accurate
measurements and safe instrument operation.
Table 1: Environment requirements
Characteristic Description
Temperature Operating instrument: 0°C to +50°C (+32°F to 120°F), with 5 °C/minute maximum gradient, noncondensing (NC)
Operating with battery: 0 °C to 45 °C (+32 °F to 113 °F)
For proper cooling, keep the rear of the instrument clear of obstructions for 2 inches (51 mm).
Table continued…