Field or control Description
Threshold Sets the amplitude level that the signal must pass through to be considered a valid transition.
Set to 50% Sets the threshold at 50% of the measured signal transition range. 50% is calculated as (Top +
Bottom)/2.
Time Limit Sets the time period condition to be met.
Copy Trigger Settings to
Search
Sets the search criteria to match the current oscilloscope trigger settings.
Copy Search Settings to
T
rigger
Sets the current oscilloscope trigger settings to match the search criteria.
Other search types
Bus Search configuration menus on page 139
Edge Search configuration menu on page 151
Logic search configuration menu on page
152
Pulse Width Search configuration menu on page 154
Rise/Fall Time Search configuration menu on page 156
Runt Search configuration menu on page 157
Setup and Hold Search configuration menu on page 158
Analog Channel configuration menu
Use the Analog Channel configuration menu to set up analog channel vertical settings, probe settings, and deskew settings for the
channel.
To open an analog Channel configuration menu, double-tap an analog Channel badge. The following text describes analog channel
settings.
Vertical Settings panel, fields and controls
Field or control Description
Display Enables (On) or disables (Off) displaying the channel waveform in the W
aveform View window.
Invert Reverses the amplitude polarity of each sample point. When On, Invert adds a down arrow symbol to
the channel badge.
Vertical Scale Sets the vertical scale using the multipurpose knob, double-tap to bring up the virtual keypad, or tap
the up and down arrows to change the scale.
Offset Sets the channel signal vertical offset.
Set to 0 Sets the channel signal vertical offset to 0 units.
Position Double-tap the field to set the vertical position using the virtual keypad.
Set to 0 Tap to set the waveform zero volt level to the center of the slice or waveform view.
Label Double-tap the field to add a label to the channel display using the virtual keypad.
Bandwidth Limit Tap to select the bandwidth limit from the drop-down list.
Coupling Tap to set the input coupling to DC or AC.
Menus and dialog boxes
2 Series MSO MSO24 and MSO22 161