Characteristic Description
Operating
humidity
5% to 90% relative humidity at temperatures up to +30°C,
5% to 60% relative humidity at temperatures greater than +30°C and up to +50°C.
Operating altitude Up to 3000 meters (9842 feet)
Battery power Requires 2-BATPK battery pack with 2 slots for batteries to be ordered with instrument or 2-BP battery pack with 2
slots for batteries to be ordered post instrument purchase
Supports up to 2 TEKBAT-XX Li-Ion rechargeable batteries. Operating time; Up to 3 hours single battery and up to 6
hours dual batteries.
Tektronix recommends using the TEKCHG-XX external battery charger to charge the TEKBAT-XX batteries in
ambient environments greater than +30°C.
Table 2: Power requirements
Characteristic Description
Power source voltage 24 V DC
Power source current 2.5 A
Input signal requirements
Keep the input signals within allowed limits to ensure the most accurate measurements and prevent damage to the analog and digital
probes or instrument.
Make sure that input signals connected to the instrument are within the following requirements.
Input Description
Analog input channels and AUX In, 1 MΩ
setting, maximum input voltage at BNC
300 V
RMS
Measurement Category II
Digital input channels, maximum input
voltage range at digital inputs
Observe probe ratings
P6316 Logic Probe
Check that the instrument passes power-on self tests
Power-on self tests verify that all instrument modules are working correctly after power up.
Procedure
1. Power on the instrument and wait until the instrument screen appears.
2. Select Utility > Self T
est from the top-edge Menu bar to open the Self Test configuration menu.
3. Check that the status of all power-on self tests are Passed.
If one or more power-on self tests shows Failed:
1. Power cycle the instrument.
2. Select Utility > Self Test. If one or more power-on self tests still shows Failed, contact Tektronix Customer Support.
Installing your instrument
18