Time gates can be placed anywhere. They can precede acquisition start value and extend beyond the acquisition value.
The Time Gating bars are shown in both the overview and zoomed view when zoom is on.
See also
Measurement Name panel (Measurement configuration menu) on page 115
Configure panel (Measurement configuration menu) on page 116
Reference Levels panel (Measurement configuration menu) on page 116
Pass/Fail Testing panel (Measurement Settings menu) on page 119
Pass/Fail Testing panel (Measurement Settings menu)
Use these settings to test measurement values against a specified limit and set actions for the instrument to take on failure.
To open the PASS/FAIL TESTING panel:
1. Double-tap a Measurement badge.
2. Tap the Pass/Fail Testing panel.
Pass/Fail Testing panel fields and controls
Not all items listed in the table may be shown for all measurements; The panel only shows fields and controls relevant to the selected
measurement.
Field or control Description
Pass/Fail Test Turns pass/fail testing on or off. Default is off.
Fail When Sets the condition for when the test on the measurement will fail. Default is < Limit.
Limit Sets the measurement limit value. This control is present when Fail When is not set to Inside Range
or Outside Range.
Low Limit Sets the low limit for the measurement value. This control is only present when Fail When is set to
Inside Range or Outside Range.
High Limit Sets the high limit for the measurement value. This control is only present when Fail When is set to
Inside Range or Outside Range.
Actions on Failure Opens a configuration menu to set the actions the instrument takes when the test fails.
See Act On Event configuration menu on page 105.
Badge Navigation Sets the navigation in the results badge.
T
ap to toggle to view Failures or All Occurrences.
When Failure is selected the navigation buttons move from failure to failure.
When All Occurrences is selected, the navigation button moves through each occurrence.
Other measurement settings panels
Measurement Name panel (Measurement configuration menu) on page 1
15
Configure panel (Measurement configuration menu) on page 116
Reference Levels panel (Measurement configuration menu) on page 116