6. For more information on serial bus settings, tap the Help button on the Bus configuration menu.
Add a parallel bus to the W
aveform view
Use this procedure to add a parallel bus to the Waveform view.
When you acquire data from a Parallel bus, you can set up the bus to be clocked or unclocked. If the bus is not clocked, the instrument
acquires all data from the parallel bus at the sample rate of the instrument.
1. Tap the Math Ref Bus > Add New Bus button on the Settings bar. This adds a Bus badge bus to the Settings bar, adds a bus
waveform to the screen, and opens the Bus configuration menu. The default bus type is parallel.
2. If setting up a clocked bus:
a. Set Clocked Data to Y
es.