The connected TekDrive is available in all file browsing windows.
Unmount a network drive from a standard instrument
Use this procedure to unmount (remove, disconnect) a network Linux mount point from a standard (non-Windows 10) instrument.
T
o unmount a network drive from your instrument:
1. Tap File > File Utilities.
2. Select the drive to unmount in the Drive column.
3. Tap Unmount. The instrument immediately unmounts the drive and removes it from the Drive column.
Unmount a TekDrive
Use this procedure to unmount (remove, disconnect) a TekDrive.
Procedure
1. Tap File > File Utilities.
2. Select the TekDrive to unmount in the Drive column.
3. Tap Unmount.
The drive is unmounted and it is removed from the Drive column.
Connect the oscilloscope to a PC using a USB cable
Use a USB cable to connect the oscilloscope directly to a PC for remote instrument control.
1. On the oscilloscope, select Utility > I/O from the menu bar.
2. Tap USB Device Port Settings.
3. Confirm that the USB Device Port control is On (default setting).
4. Connect a USB cable from the PC to the USB Device port on the instrument.
Deskew analog input channels - quick visual method
Use the following procedure to visually align waveform edges to compensate for timing differences between probes.
Critical Time measurements on multiple channels require that all probes be adjusted, or deskewed, to compensate for signal timing
differences between probes. This procedure uses displayed waveform edges to quickly minimize deskew between probes.
Note: Once probes have been deskewed for a particular channel, you should only use the probes on the channels for which they
were deskewed, when taking critical
Time measurements.
1. Connect all probes that you want to deskew
.
2. Connect up to four probe tips and ground leads to the Probe Compensation connector (maximum of four channels at a time).
3. Turn on (display on screen) the connected channels that you want to deskew.
4. Double-tap in the Waveform view and set Waveform Mode to Overlay.
5. Push the Autoset button.
6. Adjust the vertical Scale and Position controls for each channel so that the signals overlap and are centered on the display.
7. Adjust the horizontal Scale so that the differences in the channel delays are clearly visible.