Chapter 7 - Working With Measurements Measuring Vascular Exams
Terason t3000 / Echo Ultrasound System User Guide 181
Working with Vascular Reports
After making the appropriate vascular measurements, you can generate a worksheet. The
worksheet includes all stenosis measurements with text annotation and all spectral
measurements that have an anatomy. Up to three measurements can be stored per vessel or
location. If
RT or LT are not selected, the measurement is categorized as “Other”. If more
than three are saved, the first is deleted and the last is added to the worksheet. The user can
edit by checking or unchecking a measurement on the worksheet. Measurements cannot
be deleted from the report; the report is not editable.
The highest velocity of each vessel/location is displayed in the report and the highest
velocity of each vessel is used for ratios. The only exception would be if
Always use mid
CCA for ICA/CCA ratio is checked on the worksheet.
You can enter pertinent vascular history to the worksheet in the Blood pressure,
Indications and Comments fields.
The following page shows an example of a worksheet.
To produce a report you must:
1. Enter or load Patient information
2. Make the needed measurements.
3. With stenosis measurements, text annotation must be added to the image for the
measurement to appear on the report, see Working with Annotations on page 81. With
all Spectral measurements, anatomy must be applied to the measurement, see
Measuring Vessel Blood Velocities on page 171.
4. Save the image with the measurements to store them to the worksheet.