Chapter 7 - Working With Measurements Measuring in the 2D Window
Terason t3000 / Echo Ultrasound System User Guide 153
When measuring in Split Screen mode, the measurements are displayed in a single list,
regardless of which screen contains that measurement. You must make the specific screen
(left or right) active by clicking in the screen or clicking the Toggle button before you can
make a measurement.
Measuring Volume
When you measure length, width, and depth, the Terason software automatically
calculates the volume in cubic centimeters. Because you must make two scans to properly
measure volume, these instructions use the Split Screen option so you can retain both
scans.
Complete the following steps (you can make the length, width, and depth measurements in
any order that make sense for the specific scan):
1. Click . The measure tabs for the selected scan mode are added to the Image
Control window.
2. If necessary, click the 2D Measure tab or click in the 2D window.
3. Select
Length from the Measurement list.
4. Click at the starting point of the area’s length, then click at the ending point of the
area’s length.
The Terason software displays the length measurement.
5. Select
Width from the Measurement list.
6. Click at the starting point of the area’s width, then click at the ending point of the
area’s width.
The Terason software displays the width measurement.
7. Click .
The Terason software adds a new live scan adjacent to the frozen scan.
8. Click .
9. Select
Depth from the Measurement list.
10. Click at the starting point of the area’s depth, then click at the ending point of the
area’s depth.
The Terason software displays the depth measurement and the volume calculation.