Chapter 12 - System Safety and Maintenance Recommended Maintenance
Terason t3000 / Echo Ultrasound System User Guide 286
Recommended Maintenance
Terason transducers require very little maintenance. Transducers should be inspected
monthly. If the transducer is dropped, immediately examine it for signs of damage. Check
for these potential problems:
Cracks or damage to the handle, nose piece, or connector
Cuts or gouges on lens material
Swelling of lens material
Cuts or cracks in the cable
Do not expose transducers to direct heat such as strong sunlight or a local heat source, as
heat can cause crystal aging and loss of sensitivity.
When storing the Terason Ultrasound System in environments between 0º and 50ºC, the
original shipping container must be used.
All components of the Terason Ultrasound System should be checked monthly for these
potential problems:
Cuts, cracks or damage to the power cables
Dents or cracks in the electronics housing
Refer to the laptop manual for information on laptop maintenance.
Equipment List
The Terason Ultrasound System may be shipped with the following hardware:
Laptop computer with operating system and ultrasound imaging software. Refer to the
user guide that came with your computer for details about your particular computer
system.
Computer user guide
•Online Terason Ultrasound System User Guide (this document)
CD-ROMs containing the operating system and Terason ultrasound imaging software
Two (2) power cords (wall-to-transformer and transformer-to-adapter)
Computer AC/DC adapter
•ECG
ECG lead set - 10 sets of electrodes
Docking Base
•Cart
Monitor
•Printer
DVD recorder
One or more transducers:
4V2A phased-array transducer
Warning: Do not use any equipment that is cut, cracked, dented, or
displays any other damage. Use of damaged equipment can
create shock hazards for the operator and patient. Contact
Terason Customer Support for help.