3-12 Tuning the Instrument
Instrument Setup Wizard Detector Setup:
The wizard automatically calculates the detector setting that is required
to detect more than 90% of all the ions produced.
During the procedure the detector voltage is set so that the intensity of
the signal no longer increases as the detector voltage is increased. To
ensure optimum sensitivity and mass accuracy, check this value
periodically.
See also: Setting the optimum gain for the detector on page 10-52.
When completed, the Instrument Setup Wizard DDTC Setup window
appears.