Waters GCT Premier
Mass Spectrometer
Operators Guide
71500109202/Revision B
Copyright © Waters Corporation 2006.
All rights reserved.

Table of Contents

Questions and Answers

  • T
    Tina ChoiAug 21, 2025
    How to fix a Waters GCT Premier power switch that won't turn on?
    • M
      Margaret WrightAug 21, 2025
      If the power switch fails to power on your Waters Laboratory Equipment, first, unplug the instrument from the main power supply. Then, replace both mains power fuses with the correct type and rating. If the fuses continue to fail, disconnect the instrument from the power supply and contact Waters for further assistance.
  • C
    Curtis KingSep 6, 2025
    What to do if the Emission: Trap ratio is greater than 10:1 on Waters Laboratory Equipment?
    • W
      williamsjenniferSep 6, 2025
      If the Emission: Trap ratio is greater than 10:1 in your Waters Laboratory Equipment, remove the outer source magnets and reassemble them with opposite poles facing each other.
  • D
    Donna GarrisonSep 4, 2025
    What does it mean when my Waters GCT Premier Laboratory Equipment tunes with repeller more than 5 V?
    • K
      kelleystephanieSep 5, 2025
      If your Waters Laboratory Equipment tunes with repeller more than 5 V, clean the inner source.
  • N
    nicholaspowellSep 3, 2025
    What causes an inactive ion repeller in EI mode in Waters GCT Premier?
    • V
      Vicki WilliamsSep 3, 2025
      If the ion repeller is inactive in EI mode on your Waters Laboratory Equipment, use tweezers to adjust the spring contacts on the outer source to improve contact.
  • J
    jjonesSep 1, 2025
    Why does my Waters GCT Premier Laboratory Equipment have high filament current, high emission current, and low trap current?
    • R
      raymargaretSep 1, 2025
      If you are experiencing high filament current, high emission current, or low trap current with your Waters Laboratory Equipment, clean the inner source. Also, use tweezers to adjust the spring contacts on the outer source to improve contact. Check the source magnet are attracting.
  • K
    Karen CrosbyAug 31, 2025
    What causes no Filament current, Trap current maximum, or Emission current zero in Waters GCT Premier?
    • R
      rodriguezvictoriaAug 31, 2025
      If you observe no Filament current, Trap current maximum, or Emission current zero on your Waters Laboratory Equipment, remove the inner source and exit plate. Check the cleanliness of the trap and trap ceramic. Blow out any particles using dry nitrogen.
  • J
    Jeffrey HopkinsAug 29, 2025
    Why is there no Trap, Emission, or Filament current readback on my Waters GCT Premier?
    • K
      keithmoralesAug 29, 2025
      If you observe no Trap, Emission, or Filament current readback on your Waters Laboratory Equipment, remove the inner source and test the continuity of the filament with a digital multimeter (the resistance should be approximately 0.2 ohm). Replace the filament if it is damaged. If the filament is intact, use tweezers to adjust the spring contacts on the outer source to improve contact. Also, in the MassLynx Tune window, Source tab, click Reset Filament. If the solvent trip level has been exceeded, adjust the solvent trip level or correct the vacuum housing.
  • S
    Sydney CunninghamAug 27, 2025
    How do I troubleshoot a non-working source heater in Waters GCT Premier?
    • M
      Mr. Richard Fritz MDAug 27, 2025
      If your Waters Laboratory Equipment source heater is not working, inspect the Tune window EI tab Source Temp (°C) readback. A readback of 503 indicates that the source heater is disconnected. Check the integrity of the heater connections.
  • W
    whitedebraAug 25, 2025
    Why are the peaks unsteady or low intensity (ion beam) in my Waters GCT Premier?
    • M
      mandyirwinAug 25, 2025
      If you are experiencing unsteady or low-intensity peaks (ion beam) with your Waters Laboratory Equipment, consider these potential solutions: * Troubleshoot the autosampler and check for pressure variations during injection. * Clean the source components. * Ensure that the source voltage read-backs vary with the Tune window setting; if any voltages are absent, disassemble and correctly reassemble the source assembly. * Ensure all Tune window settings are correct, the readbacks are sensible, and that all lens parameters affect the beam. * Ensure the MCP voltage is set correctly.

Summary

Safety Information

Instrument Description

Preparing the Instrument for EI;CI Operation

Changing the EI;CI Inner Source Assembly

Provides instructions for removing and installing the EI/CI inner source assembly.

Changing the EI;CI Outer Source Assembly

Details the procedure for removing and fitting the EI/CI outer source assembly.

Tuning the Instrument

Preparing the Instrument for Operation

Outlines steps to prepare the instrument for operation and tuning.

Obtaining an Ion Beam

Describes the process of obtaining an ion beam for instrument tuning.

Setting Up the Instrument Automatically

Guides through automatic instrument setup using the Instrument Setup Wizard.

Tuning the Instrument for EI

Explains how to tune the instrument for Electron Ionization (EI) mode.

Tuning the Instrument for Best Resolution in EI

Details procedures to achieve the best resolution in EI mode.

Tuning the Instrument for Positive CI

Explains how to tune the instrument for positive Chemical Ionization (CI).

Tuning the Instrument for Negative CI

Guides on tuning the instrument for negative Chemical Ionization (CI).

Calibration

Calibrating the Instrument

Outlines the process of calibrating the instrument's mass scale.

Using the Calibration Wizard

Guides through the instrument calibration process using a wizard.

Calibrating the Instrument in CI

Explains how to calibrate the instrument in Chemical Ionization (CI) mode.

Manual Calibration

Provides instructions for manual instrument calibration procedures.

The GC Interface

Installing the GC Interface

Provides step-by-step instructions for installing the GC interface.

The Direct Insertion Probe

Inserting and Removing the DI Probe

Details tasks for loading, inserting, and removing the DI probe.

Controlling the DI Probe Temperature

Explains how to control the DI probe temperature manually and automatically.

The Direct Chemical Ionization (DCI) Probe

Installing the Probe Insertion Lock

Guides on installing the probe insertion lock for the DCI probe.

Inserting and Removing the DCI Probe

Details tasks for loading, inserting, and removing the DCI probe.

Controlling the DCI Probe Current

Explains how to manually control the DCI probe current.

FI;FD Operation

Setting-up for FI;FD Operation

Guides on setting up the instrument for FI/FD operation.

Tuning in FI Mode

Explains how to tune the instrument in Field Ionization (FI) mode.

Setting Up and Calibrating the Instrument for FI

Guides on setting up and calibrating the instrument for FI operation.

Tuning in FD Mode

Explains how to tune the instrument in Field Desorption (FD) mode.

Routine Maintenance

Maintenance Schedule and Safety

Lists periodic maintenance schedules and safety precautions.

Rotary Pump Oil and Maintenance

Discusses rotary pump oil quality, color, gas ballast, and maintenance.

EI;CI Inner Source Maintenance

Covers cleaning and maintenance of EI/CI inner source components.

FI;FD Inner Source Maintenance

Details maintenance procedures for FI/FD inner source, including emitter care.

EI;CI Outer Source Maintenance

Covers cleaning and maintenance of EI/CI outer source components.

FI;FD Outer Source Maintenance

Details maintenance procedures for FI/FD outer source components.

Checking for Leaks after Source Maintenance

Guides on checking for leaks after performing source maintenance.

Detector Maintenance

Covers maintenance procedures for the detector, including conditioning.

Troubleshooting

Component Hardware Troubleshooting

Provides tables of common hardware problems and corrective actions.

Contacting Waters for Support

Gives information on how to contact Waters for technical support.

Starting-up and Shutting-down the Instrument

The MassLynx Tune Window

Waters GCT Premier Specifications

General IconGeneral
SoftwareMassLynx
Resolution10, 000 FWHM
Ionization ModesEI, CI
Mass Accuracy<5 ppm
Dynamic Range>4 orders of magnitude