5-36
IM 760201-01E
Setting Operand Parameters
The parameters that you need to enter depend on whether the function is followed by “( , )”
or “( )”.
Setting parameters for functions followed by “( , )”
Specify the element to the left of the comma, and specify the harmonic order to the
right of the comma. For example: (E1,OR2).
Symbols used to represent elements
E1 to E3: Element 1 to element 3
E4:
Σ
Symbols used to represent harmonic orders
*
ORT: Total
OR0: DC
OR1: Fundamental signal
OR2 to OR50: Harmonics
* The harmonic measurement feature (optional) is required.
Setting parameters for functions followed by “( )”
Specify the element. You do not need to specify a harmonic order. For example: (E1).
For information about the symbols that can be used in the parameters of each operand,
see appendix 7.
Values Substituted in Operands
The substituted value of TI() is the same regardless of which element symbol (E1 to
E3) is used in TI(). The unit of TI values is seconds.
The U1 in PHIU1U2 represents the voltage signal of the element whose element
number is the smallest in the wiring unit (
Σ
). For example, if input elements 2 and 3
are assigned to wiring unit
Σ
, PHIU1U2 is the phase difference between the voltage
signals of input elements 2 and 3.
5.13 Setting User-Defined Functions