4-16
IM 760201-01E
4.6 Setting the Scaling Feature When Using a VT
or CT
Procedure
1.
Press SETUP to display the Setup menu.
2.
Use the cursor keys to select Scaling.
3.
Press SET to display the Scaling Settings menu.
Turning the Scaling Feature On or Off
4.
Use the cursor keys to select Scaling.
5.
Press SET to select ON or OFF. If you select ON, the Scaling indicator lights at
the top of the screen (see section 1.3 for details).
Setting the VT Ratio
6.
Use the cursor keys to select the VT of the element that you want to set the VT
ratio of.
7.
Press SET to display the VT ratio entry box.
8.
Use the cursor keys to set the VT ratio.
9.
Press SET or ESC to close the VT ratio entry box.
10.
Repeat steps 6 to 10 to enter the VT ratios for all of the elements.