14.9 USB PC Interface
Item Specifications
Connector Type B connector (receptacle)
Electrical and mechanical specifications USB 1.1
Data rate 12 Mbps max.
Ports 1
Power supply Self powered
Supported services Remote control
System requirements A
PC running Windows 2000 (SP4 or later), Windows XP, or Windows Vista and USB
ports.
14.10 USB PERIPHERAL Interface
Item Specifications
Connector type USB type A (receptacle)
Electrical and mechanical specifications USB 2.0
Data rate 480 Mbps max.
Supported keyboards
1
104 (US) or 109 (Japanese) keyboards compliant with USB HID Class Version 1.1
Supported USB mass storage USB flash memory (that meets the USB Mass Storage Class specifications)
Power supply 5 V, 500 mA
2
(for each port)
Ports 2
1
For more detailed information about supported USB devices, contact your nearest YOKOGAWA dealer.
2 You cannot connect devices whose maximum current consumption exceeds 100 mA to two different ports on the WT500 at the
same time.
14.11 GP-IB Interface Option
Item Specifications
Usable devices NATIONAL INSTRUMENTS
AT-GPIB
PCI-GPIB or PCI-GPIB+
PCMCIA-GPIB, PCMCIA-GPIB+, or PCIe-GPIB
Driver: NI-488.2M version 1.60 or later
Electrical and mechanical specifications IEEE St’d 488-1978 (JIS C 1901-1987)
Functional specifications SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, and C0
Protocol IEEE St’d 488.2-1992
Code ISO (ASCII)
Addresses 0 to 30
Mode Addressable mode
Clearing remote mode The LOCAL key can be pressed to clear remote mode (except during Local Lockout).