6-19
IM 760201-01E
Harmonic Measurement (Optional)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Setting the Bar Graph Display Range
The bar graph display range is set in terms of harmonic orders.
The display ranges of Graph1 to Graph3 are the same.
The minimum harmonic order is 0 (DC). However, if the measurement function is set
to
f
,
f
U, or
f
I, the 0
th
order has no value. Thus, the 0
th
order is not displayed in the
bar graph.
If the measurement function is set to
f
U or
f
I, the 1
st
order has no value. Thus, the 1
st
order is not displayed in the bar graph.
The maximum settable harmonic order is 50. However, bar graphs of orders that
exceed the upper limit of harmonic analysis (see section 6.4 for details) are not
displayed.
Cursor Measurement
ON/OFF
A cursor can be placed on the displayed bar graph to measure the value at that point.
ON: Cursor measurement is performed.
OFF: Cursor measurement is not performed.
Measured Items
Y1+: Y-axis value of cursor + of bar graph 1
Y1x: Y-axis value of cursor x of bar graph 1
Δ
Y1:
Difference between the Y-axis values of cursor + and cursor x of bar graph 1
Y2+: Y-axis value of cursor + of bar graph 2
Y2x: Y-axis value of cursor x of bar graph 2
Δ
Y2:
Difference between the Y-axis values of cursor + and cursor x of bar graph 2
Y3+: Y-axis value of cursor + of bar graph 3
Y3x: Y-axis value of cursor x of bar graph 3
Δ
Y3:
Difference between the Y-axis values of cursor + and cursor x of bar graph 3
Moving the Cursors
Two cursors (+ and x) are displayed in each graph (Graph1 to Graph3).
The cursor positions are set by harmonic order.
The bar graph display indicates what harmonic order each cursor is located in. For
example:
The location of cursor + is indicated in this format: “order+:2.”
The location of cursor x is indicated in this format: “orderx:45.”
Cursors + and x are located in the same harmonic orders in each graph (Graph1 to
Graph3).
Note
If immeasurable data exists, “***” is displayed in the measured value display area.