13-4
IM 760201-01E
Error in Setting
Code Message Solution Refer to
section
800 Illegal date-time. Set the correct date and time. 3.12
801 Illegal file name. The file name contains characters which are not
allowed or the file name is not a valid MS-DOS file
name. Enter another file name.
10
808 Cannot change this parameter during
numerical display.
809 Cannot change this parameter during
vector display.
810 Cannot change this parameter during
bar graph display.
812 Cannot be set while storing data.
813 Cannot be set while integration is running. Reset Integration. 5.9
823 Cannot change during CAL. Wait until CAL is completed. 12.3
827 Illegal math expression. Input a correct computing equation. 5.4
831 Processing now. Retry setting or execution again.
841 Attempted to start integration after integration
time has reached its preset value.
Reset Integration. 5.9
842 Attempted to start integration while integration
is in progress.
5.9
843 Measurement stopped due to overflow during
integration or due to a power failure.
Reset Integration. 5.9
844 Attempted to stop integration even though
integration was not in progress.
5.9
845 Attempted to reset integration even though
integration was in progress or integration
mode was not selected.
5.9
846 Attempted to start integration while measurement
of peak overflow was in progress.
5.9
847 Attempted to start integration in continuous
integration mode when integration preset time
was set to “0”.
5.8
848 Attempted made to start integration in real time
counting integration mode when the end time
had already passed.
5.9
849 Attempted made to start storing in real time
counting storing mode when the end time
had already passed.
9.3
850 Cannot be set or executed at current store state.
To set or execute, reset srore.
9.4
852 Stored file is illegal.
Initialize memory before storing.
9.4
854 Waveform display data not found.
855 Store data destination memory is full.
Storing has been stopped.
9.4
856 An error has occured while storing.
Storing has been stopped.
857 Cannot be set while Master/Slave Synchronized
Measurement is set to Slave.
12.6
858 Store process is in progress now. Execute or set setting again. 9.4, 9.6
859 Cannot convert selected file. Select a file with an extension of WTS or HDS. 9.7
862 Numeric data not found.
864 This wiring cannot be set as the first selected
element.
4.2
865 Cannot be set while integration is running. Stop or reset Integration. 5.7
873 When 500 Hz is selected as the line filter,
the freq filter is set ON and can not be changed.
PLL source or trigger source cannot be set to Ext
Clk, while Master/Slave
To change the freq filter, choose a line filter other
than 500 Hz.
4.8
874 Sync source, PLL source or trigger source cannot be
set to Ext Clk, while Master/Slave Synchronization
Measurement is set to Slave.
4.7, 6.3,
7.5, 12.6
13.2 Error Messages and Troubleshooting Methods