Item Specifications
Line filter influence When the cutoff frequency is 500 Hz
45 to 66 Hz: Add 0.3% of reading. Less than 45 Hz: Add 1% of reading.
When the cutoff frequency is 5.5 kHz
66 Hz or less: Add 0.4% of reading. 66 to 500 Hz: Add 1.2% of reading.
Temperature coefficient Same as the temperature coefficients for the voltage and current.
One year accuracy Reading error of accuracy at 6 months × 1.5.
Accuracy when the crest
factor is set to 6
Same as the accuracy when the crest factor is 3 after doubling the measurement range.
Other Computation and Measurement Accuracies
Accuracy of apparent power S Voltage accuracy + current accuracy
Accuracy of reactive power Q
Accuracy of apparent power + (
(1.0004 L
2
)
(1 L
2
)
) ×
100% of range
Accuracy of power factor
λ
±[(
λ
λ
/1.0002) + |cos
f
– cos{
f
+ sin – 1(influence from the
power factor when
λ
= 0%/100)}|] ± 1 digit
when voltage and current are at the measurement range rated
input.
Accuracy of phase difference
f
±[|
f
– cos – 1(
λ
/1.0002)| + sin – 1{(influence from the power
factor when
λ
= 0%)/100}] deg ± 1 digit
when voltage and current are at the measurement range rated
input.
For the accuracies of frequencies fU and fI, see “Frequency Measurement” in section 14.6,
“Functions.”
For the accuracies of integrated values WP, WP+, WP–, q, q+, and q– and the integration time
(Time), see “Integration” in section 14.6, “Functions.”
For the accuracies of voltage, current, and power during harmonic measurement, see “Harmonic
Measurement” in section 14.6, “Functions.”
14.5 Accuracy