TECHNICAL MANUAL
Axio Scope.A1 Technical Data Carl Zeiss
M60-2-0007 e 05/08 27
2.3 Technical Data
Dimensions (width x depth x height)
Microscope stand Axio Scope.A1 (no DL illumination) ..................... approx. 240 mm x 340 mm x 365 mm
Stand column Vario 380 mm .......................................................... approx. 460 mm x 390 mm x 465 mm
Stand column Vario 560 mm .......................................................... approx. 460 mm x 390 mm x 645 mm
Mass
Microscope stand Axio Scope.A1 (depending on variant and accessories).................... approx. 14 to 20 kg
Stand column Vario ............................................................................................................. approx. 19 kg
Environmental conditions
Transport (in packaging):
Permissible environment temperature............................................................................. -40 to +70 °C
Storage:
Permissible environment temperature............................................................................. +10 to +40 °C
Permissible humidity (no condensation).................................................................. max. 75 % at 35 °C
Operation:
Permissible environment temperature.............................................................................. -10 to +40 °C
Permissible relative humidity (no condensation)...................................................... max. 75 % at 35 °C
Height of application........................................................................................................ max. 2000 m
Air pressure.......................................................................................................... 800 hPa to 1060 hPa
Pollution level......................................................................................................................................2
Operational data
Range of application........................................................................................................................indoor
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical security ................................................................... according to DIN EN 61010-1 (IEC 61010-1)
...............................................................................................In consideration of CSA and UL regulations
Overvoltage category............................................................................................................................... II
Radio interference suppression ................................................................... according to EN 55011 class B
Immunity.........................................................................................................according to DIN EN 61326
Voltage range.....................................................100 to 240 V ±10 %, no voltage adjustment is necessary no vo
Power frequency..........................................................................................................................50/60 Hz
Power input Axio Scope.A1 with internal power supply unit ......................................................... 110 VA
Power input Axio Scope.A1 with auxiliary power supply unit 12 V DC 100 W ............................... 220 VA
Transformer mbq52ac-z for HBO 50
Range of application........................................................................................................................indoor
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Switchable voltage range....................................................... 100, 110, 120, 127 VAC and 230, 240 VAC
Switchable power frequency................................................................................................. 50 and 60 Hz
Power input with HBO 50 in operation ..................................................................................max. 350 VA