OPERATION
Axio Scope.A1 Illumination and Contrasting Method Carl Zeiss
M60-2-0007 e 05/08 73
Replace the eyepiece into the tube.
Lighter areas which might still be visible in the
field of view are reduced when the dark-field
condenser is adjusted correctly and delicately.
You now have an approximately sharp eye
diaphragm image.
Now adjust the field diaphragm diameter to the
size of the field of view.
Dark-field samples need to be kept impeccably
clean, more so than samples for any other method.
A fingerprint, dust or any dirt particle can have a
negative effect, as they brighten the background
and reduce the contrast of the object image.
Fig. 4-4 Centering the dark-field diaphragm
on the condenser, achromatic-
aplanatic 0.9 H D Ph DIC