EasyManua.ls Logo

Zeiss Axio Scope.A1 - Page 94

Zeiss Axio Scope.A1
123 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATION
Carl Zeiss Illumination and Contrasting Method Axio Scope.A1
94 M60-2-0007 e 05/08
Switch on the lamp HBO 100 or HBO 50
(
691H691HFig. 4-16/1) and let it warm up to operational
temperature for about 15 min.
On the reflector turret/slider (
692H692HFig. 4-16/6),
choose the reflector module FL P&C with the
desired fluorescence filter combination
(depending on the excitation mode) and switch
it on.
Now unblock the light path in the reflected
light illuminator with the filter slider reflected
light (
693H693HFig. 4-16/2).
If necessary, adjust FL attenuator to 100 %
transmission in order to facilitate locating
fluorescence signals. Reduce the transmission
later on in order to preserve the sample.
Take out an eyepiece from the tube and adjust
the aperture diaphragm at sight. Open the
compartment for the aperture diaphragm for
P&C (
694H694HFig. 4-16/3) enough to see the whole
objective exit pupil.
Replace the eyepiece in the tube and close the
field diaphragm (
695H695HFig. 4-16/4) enough to make it
visible in the field of view.
Using both centering screws (
696H696HFig. 4-16/5), center the field diaphragm on the edge of the field of view.
Open the field diaphragm enough to make it just disappear behind the edge of the field of view, or, if
you are using a sample which might bleach out, reduce the field diaphragm into the field of view.
Focus the sample again and optimize the collector position of the HBO 100. Adjust the collector so
that the reflector module of the short-wave excitation illuminates the field of view consistently. A
correction of the collector position is not necessary in modules with longer-wave excitation.
Fig. 4-16 Components for reflected light
fluorescence

Table of Contents

Related product manuals