Precauons and Limitaons
For best results, observe the following precauons when using the i‑STAT Alinity printer.
DO:
• use only a rechargeable baery pack purchased from Abbo Point of Care.
• use only the power adapter and supply provided with the i‑STAT Alinity printer kit.
• use an i‑STAT Alinity printer when aempng to print from an i‑STAT Alinity instrument.
• check with authories for local, state, and/or naonal requirements for disposal.
DO NOT:
• operate the printer without paper.
• pull paper through the paper mechanism. Damage to the printer can result. Use the paper feed
buon.
• allow the power supply to become a trip hazard.
• place the printer in an oxygen-enriched atmosphere.
• disturb the instrument or the printer unl prinng is complete, since this will interrupt the printout. If
prinng is interrupted, realign the instrument and the printer, or replace the instrument in the Base
Staon to resume prinng.
• place printers side by side. Doing so may cause instrument to printer communicaon problems.
NOTE:
• The instrument and its peripherals are not listed by any authority with respect to suitability for use in
oxygen-enriched atmospheres.
• In the event of a baery leaking, do not allow any leakage to come into contact with the skin or eyes.
If contact has been made, wash the aected area with copious amounts of water and seek medical
advice.
• If a printout appears inconsistent with a paent's clinical assessment, verify that the correct paent
record was selected (paent ID, date and me of test, etc.). If the record is not correct, select the
correct record and print. If the printout sll does not match the data in the instrument, the printer
needs service and the printed results must not be used. If another printer is available, retry.
• Fluorescent light sources can cause interference with communicaons sent to the i‑STAT Alinity
printer. When light from a uorescent source of sucient proximity or brightness has a direct path
into the printer's infrared radiaon (IR) window, the printer might fail to respond when records are
sent for prinng over a serial (wired) connecon to a Base Staon.