Cal Ver Schedules
This feature is used to dene a Cal Ver tesng schedule for the i‑STAT Alinity instruments. A customized
Cal Ver tesng schedule can include: the cartridge types and Cal Ver uids to be used, the cartridge types
that are enabled by running the Cal Ver uids, and the schedule for performing Cal Ver tesng. Up to 3
Cal Ver tesng schedules can be dened. The opons for schedules are shown here once, since the
opons are idencal for all three.
Note: Not all products are available in all regions. Check with your local representave for
availability in specic markets.
Frequency Select one of the following:
O This is the default.
Every_____ ______ of
each month
The rst blank is lled in with rst, second, third or
fourth chosen from drop down list. The second
blank is lled in with day of the week chosen from
the drop down list. Example - Every 2nd Thursday
of the month.
Time This opon is enabled only when a value other
than O has been specied for Frequency.
Tesng due at HH:mm Time of day when Cal Ver is due to be done.
Specify a value for hours and minutes in the range
00:00 to 23:59. Default is 00:00 (midnight).
Grace period Addional amount of me allowed before and
aer tesng due me to perform Cal Ver tesng if
tesng is not done at the scheduled me. Aer
this amount of me elapses, paent tesng is
disabled. Grace period also determines when the
alert message for scheduled Cal Ver will display on
the instrument.
n hours
Number of hours. 8 hours is default. Range is 0 to
255.
Apply Schedule to Select which months to perform tesng. Any or all
months of the year can be selected.
Cartridge QC Prole If Frequency is specied, dene at least one
Cartridge QC Prole.
Add QC Prole Click Add QC Prole. This panel displays:
201
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022