EasyManua.ls Logo

Abbott i-STAT Alinity - 1.1 - Complete i-STAT Alinity System Overview; 1.3 - i-STAT Alinity Results Screen

Abbott i-STAT Alinity
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1.1 - Complete iSTAT Alinity System Overview
The iSTAT Alinity instrument is an analycal, in vitro diagnosc device. The instrument requires iSTAT
single-use cartridges containing electrodes and sensors to perform quantave diagnosc tesng on
whole blood or plasma. Together, the instrument and cartridge allow the user to perform clinical tesng
and related administrave tasks.
The iSTAT Alinity design enables the instrument to be taken to the paents bedside (point of care), a
convenient locaon near the point of care, or clinical laboratory seng.
Aer the inseron of a lled test cartridge, the instrument carefully monitors and controls the tesng
process. The only user intervenon is in the form of data entry, performed via the touchscreen or by
barcode capture. Throughout the cycle, the instrument performs a series of quality checks. These checks
are designed to monitor the status of the instrument and the quality of the cartridge. An iSTAT Alinity
instrument, a cartridge with the required test, and two or three drops of blood will allow the caregiver to
view quantave test results within minutes.
For the purposes of this system operaons manual, the iSTAT Alinity system components and associated
accessories will be discussed. Addional informaon related to the enre system can be found in the
iSTAT Alinity system documentaon listed below.
Note: Note Regarding System Reliability: The i-STAT System automacally runs a
comprehensive set of quality checks of analyzer and cartridge performance each me a
sample is tested. This internal quality system will suppress results if the analyzer or
cartridge does not meet certain internal specicaons (see Quality Control secon in
System Operaons Manual for detailed informaon). To minimize the probabilty of
delivering a result with medically signicant error, the internal specicaons are very
stringent. It is typical for the system to suppress a very small percentage of results in
normal operaon given the stringency of these specicaons. If however, the analyzer or
cartridges have been compromised, results may be persistently suppressed, and one or
the other must be replaced to restore normal operang condions. Where unavailability
of results while awaing replacement of analyzers or cartridges is unacceptable, APOC
recommends maintaining both a back up i-STAT System instrument and cartridges from
an alternate lot number.
i
STAT Alinity Documentaon:
iSTAT Alinity System Operaons Manual, including:
iSTAT Alinity Reference
i-STAT Cartridge Instrucons for Use
AlinIQ NCi - Network Connecvity for iSTAT Alinity
AlinIQ CWi - Customizaon Workspace for iSTAT Alinity
iSTAT Alinity Quick Reference Guide
iSTAT Alinity Geng Started Guides:
iSTAT Alinity Base Staon
iSTAT Alinity Rechargeable Baery
iSTAT Alinity Electronic Simulator
iSTAT Alinity Printer
3
i-STAT Alinity — System Operaons Manual Art: 745526-01 Rev. H Rev. Date: 04-Mar-2021

Table of Contents

Other manuals for Abbott i-STAT Alinity

Related product manuals