Calibraon Vericaon tesng and i-STAT TriControls
Calibraon Vericaon Soluons
A ve-level calibraon vericaon set is available to verify the calibraon of i‑STAT cartridges throughout
the reportable range.
i-STAT Calibraon Vericaon (Cal Ver) Soluons
Sodium PCO
2
Glucose
Potassium PO
2
Lactate
Chloride pH BUN/Urea
Ionized Calcium Creanine
i-STAT TriControls Calibraon Vericaon Soluons
Sodium PCO
2
Glucose
Potassium PO
2
Lactate
Chloride TCO
2
BUN/Urea
Ionized Calcium Hematocrit Creanine
pH
Each set contains four 1.7 mL glass ampules of each level.
St
orage
Store refrigerated at 2 to 8°C (35 to 46°F) unl the printed expiraon date on the box and ampule labels.
Unopened ampules of i‑STAT Calibraon Vericaon and i‑STAT TriControls Calibraon Vericaon and
soluons can also be maintained at room temperature (18 to 30°C or 64 to 86°F ) for up to 5 days.
Do not use i‑STAT Calibraon Vericaon and i‑STAT TriControls Calibraon Vericaon soluons past
the labeled expiraon date on the box and ampule labels.
Ampule use
Barcode on the ampule must be scanned before the ampule is opened.
i‑STAT Calibraon Vericaon and i‑STAT TriControls Calibraon Vericaon ampule tesng must be
performed within 10 minutes of opening the ampule.
i‑STAT Calibraon Vericaon and i‑STAT TriControls Calibraon Vericaon ampule tesng must be
performed immediately upon opening the ampule when tesng pH, PCO
2
, PO
2
, iCa.
Ranges
Assigned ranges are provided in the form of value assignment sheets (VAS). These are located on hps://
www.globalpointofcare.abbo/en/support/istat-brand/istat-alinity.html. Abbo Point of Care makes