available both eVAS and ReVAS. ReVAS ranges are provided as a feature for the German market and for
those customers who prefer ranges set using Rilibak guidelines.
Electronic value assignment sheets (eVAS or ReVAS) may be downloaded to a customized instrument via
the Abbo Managed Server, the SDi, or USB. See the System Operaons Manual for informaon and
instrucon on creang customizaon proles. Selecon of eVAS or ReVAS must be made during the
creaon of the prole.
Note: Follow facility policy regarding control results that do not fall within assigned
ranges.
Disposal
Check with authories for local, state, and/or naonal requirements for disposal.
Procedure for Tesng
Prerequisites
• Ampules, cartridges, and instruments must be at the same temperature.
• i‑STAT Calibraon Vericaon and i‑STAT TriControls Calibraon Vericaon soluons require
dierent temperature stabilizaon mes depending on whether or not PO
2
is to be measured. If PO
2
is to be measured, equilibrate the ampule for 4 hours. If not, equilibrate the ampule for
approximately 30 minutes at room (ambient) temperature.
• Do not use the soluon le in a syringe, ampule or capillary tube for addional tesng of cartridges
that contain sensors for iCa, pH, PCO
2
, or PO
2
. However, cartridges without these sensors may be
tested with remaining uids if within 10 minutes of opening the ampule.
• Since aqueous based soluons such as controls lack the buering capabilies of whole blood, the
transfer process from ampule to cartridge must be more expedient than with a paent sample.
• Refer to the value assignment sheets posted on the Abbo Point of Care website at
www.globalpointofcare.abbo.
Procedure
1. Press Power and allow the instrument to power on.
Note: Aer powering on the instrument, one or more alert messages may display.
Read the message carefully and perform the funcons necessary to evaluate and/or
clear the alert. The Home screen will display when the alerts have been successfully
managed.
2. From the Home screen touch More Opons > Quality Opons > Cal/Ver. Select the buon that is
appropriate for the tesng.
3. Enter the required informaon by following the prompts on the screen. Prepare the cartridge and
control soluon for tesng.
4. Shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To shake,
hold the ampule at the p and boom with forenger and thumb to minimize increasing the
temperature of the soluon. If necessary, tap the p of the ampule to send soluon back into the
boom secon of the ampule.
5. Protect ngers with gauze, ssue or glove, or use an ampule breaker to snap o the p of the ampule
at the neck.
6. Immediately transfer the soluon from the ampule into a capillary tube or syringe (refer to
informaon for transferring the control soluon, below), and then immediately transfer the soluon
81
i-STAT Alinity — System Operaons Manual Art: 745537-01 Rev. F Rev. Date: 02-Nov-2022