Repeat Test Item
Use the Repeat Test Item to prompt the operator to answer "Yes" or "No" if a repeat test is needed.
On the screen that displays, enter values for:
Item Name Required. Specify a name that is from 1 to 20 characters long and:
• does not begin with the characters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon.
Item Prompt on Analyzer Required. Specify a prompt that is from 1 to 20 characters long and:
• does not begin with the char
acters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon.
Descripon Oponal.
Entry Select one of the following:
Entry, oponal
Pr
ompt the operator for an oponal entry.
Entry, mandatory with override
Prompt the operator for a mandatory entry. If the override feature
is chosen by the operator no further acon is required and the
instrument will advance to the next screen.
Entry, mandatory
Prompt the operator for a mandatory entry. Cartridge tesng
cannot connue unl informaon is entered.
Under Selecon Lis
t the values Yes and No are shown. These are the only possible responses displayed
on the instrument.
• Click Publish to complete the item so it will appear on the Manage Items list
• Click Discard to delete the item
237
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022