Precauons and Limitaons
For safety when using the Base Staon:
DO:
use only the accessories and consumables specied or supplied for this system by Abbo Point of
Care.
use only the power supply shipped in the box with the Base Staon. Aempng to use a dierent
type of manufacturer's adapter could damage the unit and cause re or explosion hazards.
be sure to install all cables and power supplies so they do not pose a trip hazard. Mount equipment
so cables and accessories stay clear of walkways.
always use the Base Staon for charging. Refer to the rechargeable baery Geng Started Guide on
instrucons for proper charging.
connect only Abbo Point of Care provided printers to the Base Staon printer port.
check with authories for local, state, and/or naonal requirements for disposal.
DO NOT:
try to connect any non-electrically isolated equipment to the Base Staon.
connect the Base Staon to unauthorized medical devices or other equipment.
place metal objects on or near the exposed baery charging contacts.
place the instrument in an oxygen-enriched atmosphere.
make any unauthorized repairs or modicaons to this product, as it may cause personal injury or
damage to the unit.
NOTE:
The AC adapter power supply plug acts as a disconnect device for the Base Staon and therefore, the
AC socket outlet must be installed (or located) near the Base Staon and must be easily accessible if
the Base Staon needs to be unplugged. When correctly connected to power, the blue light on the
Base Staon will be illuminated. Aer an instrument is docked, the blue light above the screen will be
illuminated. This may take several seconds. If the blue light does not illuminate, check the power
supply. Ensure that the Base Staon is connected properly and that the instrument is seated correctly
in the Base Staon.
Protecon provided by this equipment may be impaired if used in a manner not specied by Abbo
Point of Care.
The instrument and its peripherals are not listed by any authority with respect to suitability for use in
oxygen-enriched atmospheres.
The Base Staon must be on a at, level, horizontal surface such as a table top. Falling equipment
may cause injury. Always place the equipment on a stable level surface in a locaon where it will not
cause injury if dropped.
53
i-STAT Alinity — System Operaons Manual Art: 745528-01 Rev. G Rev. Date: 16-May-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals