EasyManuals Logo

Abbott i-STAT Alinity User Manual

Abbott i-STAT Alinity
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #232 background imageLoading...
Page #232 background image
Electronic Simulator
Tesng for the external Electronic Simulator can be scheduled to be performed based on an interval of a
number of days. This seng is usually determined by compliance with regulatory agencies.
Note: Internal Electronic Simulator tesng is performed every me a cartridge is tested
on the iSTAT Alinity.
Simulator Tesng Schedule
Run Electronic Simulator every n
days
Determines how frequently the external electronic simulator is
tested. Specify a number of days in the range 0 - 365. The default is
0, which turns o tesng.
Note: Abbo Point of Care manufacturer's requirement is
for the electronic simulator to be tested every 6 months
(182 days).
When nished, choose from the opons shown at the boom of the screen:
Previous to go back to the previous screen
Discard to undo changes
Finish Later to create a dra
Next to open the next tab
192
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022