Acon Range Comment
The Acon Range Comment item prompts the operator to select a comment for any paent result that
is within the acon range.
Item Name Required. Specify a name that is from 1 to 20 characters long and:
does not begin with the characters APOC (in upper or lower
case)
is a unique name within this healthcare organizaon.
Item Prompt on Analyzer Required. Specify a prompt that is from 1 to 20 characters long and:
does not begin with the char
acters APOC (in upper or lower
case)
is a unique name within this healthcare organizaon.
Descripon Oponal.
Entry Prompt Select one of the following:
Entry, oponal
Pr
ompt the operator for an oponal entry.
Entry, mandatory with override
Pr
ompt the operator for a mandatory entry. If the override feature
is chosen by the operator no further acon is required and the
instrument will advance to the next screen.
Entry, mandatory
Pr
ompt the operator for a mandatory entry. Cartridge tesng
cannot connue unl informaon is entered.
Each item for the selecon lis
t is dened next.
Click Edit Selecon List. A text box displays:
Enter up to 20 characters in the text box. This will be the r
st item in the selecon list. To create
addional items, click Add Item.
238
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022