Aer all selecon list items have been created, click Update Selecon List or Cancel.
• Click Publish to complete the item so it will appear on the Manage Items list
• Click Discard to delete the item
QC Auto Fail Comment
The Result Notes QC Auto Fail Comment is only applied when QC Pass/Fail determinaon is set to Auto
via eVAS (refer to Quality Sengs Category).
Item Name Required. Specify a name that is from 1 to 20 characters long and:
• does not begin with the char
acters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon.
Item Prompt on Analyzer Required. Specify a prompt that is from 1 to 20 characters long and:
• does not begin with the char
acters APOC (in upper or lower
case)
• is a unique name within this healthcare organizaon.
Descripon Oponal.
Entry Prompt Select one of the following:
Entry, oponal
Pr
ompt the operator for an oponal entry.
Entry, mandatory with override
Prompt the operator for a mandatory entry. If the override feature
is chosen by the operator no further acon is required and the
instrument will advance to the next screen.
Entry, mandatory
Prompt the operator for a mandatory entry. Cartridge tesng
cannot connue unl informaon is entered.
Each item for the selecon lis
t is dened next.
Click Edit Selecon List. A text box displays:
239
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022