Specicaons
i-STAT Alinity Base Staon Specicaons
Dimensions: Height x Width x Depth 11.8 in (29.9 cm) x 5.9 in (15 cm) x 2.7 in (6.9 cm)
without AC adapter
Weight 2.4 lb (1090 g)
Communicaon Interface Ethernet 10/100 base t, RS-232, USB 2.0
LED Indicators Color: Blue
Status: Power
Operaonal range: Temperature and
humidity
10 to 40ºC (50 to 104ºF)
10 to 90% non-condensing, with maximum saturaon
temperature of 34ºC (93.2F)
Tesng Environment Indoors, on a dry, clean, horizontal, stable surface.
Avoid nearby vibraon equipment such as centrifuges.
Avoid direct sunlight.
Altude up to 3,048 meters (10,000 feet)
Storage range: Temperature and humidity -10 to 60ºC (14 to 140ºF)
10 to 90% non-condensing, with maximum saturaon
temperature of 50ºC (122ºF)
External Power Supply Unit Input: 110-240 VAC, 50-60 Hz, 1.5A
Output: 5.3V DC, 6.6A