Issue Soluon
Category Name or Prole Name displays an error
message
The following characters are not allowed when
naming categories or proles:
< (less than)
: (colon)
/ (forward slash)
| (vercal bar or pipe)
* (asterisk)
> (greater than)
" (double quote)
\ (backslash)
? (queson mark)
Forgot AlinIQ CWi Password
Launch CWi and enter Email address. Click the
Forgot Password? link. Check your email for a
message received from
pointofcare_services@noreply.abbo.com Follow
the instrucons in the message to reset your
password.
No sound on i‑STAT Alinity Check General Sengs Category to conrm that
sound is enabled on the prole assigned to the
instrument. Follow CWi instrucons in this manual
to make necessary changes.
Date display is incorrect on i‑STAT Alinity Check General Sengs Category for Date Display.
Follow CWi instrucons in this manual to make
necessary changes.
i‑STAT Alinity wireless funcon does not work Check General Sengs Category to conrm the
check box has a check enabling the Wireless
Communicaon. Check the NCi sengs for
wireless connecvity.
Operator barcode will not scan Check Barcode sengs in the Operator Sengs ID
Entry feature on the prole assigned to the
instrument. Follow CWi instrucons in this manual
to change the barcode type.