Hematocrit
On this tab, select sengs for tesng Hematocrit.
Hematocrit Seng (Calibraon by
Laboratory Hematology Analyzer)
Specify the ancoagulant used to calculate hematocrit result. For
best agreement of iSTAT Alinity and hematology analyzer
hematocrit results, the iSTAT Alinity customizaon seng is
selected according to the calibraon of the comparave
hematology analyzer (MHK
2
EDTA or MHK
3
EDTA). Since most
clinical hematology analyzers are calibrated by the
microhematocrit method using K
3
EDTA ancoagulant, the iSTAT
Alinity System default customizaon is K
3
EDTA.
Select either:
K2EDTA
K3EDTA
The default is K3EDTA.
Apply CPB Protein Algorithm to
Hema
tocrit
The Apply CPB opon is intended for use when samples are
collected from paents on cardiopulmonary bypass. The CPB
funcon adjusts hematocrit and hemoglobin results for the
diluonal eect of pump uid during cardiopulmonary bypass
surgery. However, the facility may validate its use for other
paent populaons known to have protein levels signicantly
lower than the normal adult populaon.
Note: If an ins
trument customized as “CPB, always
apply” is used for paents who are not on the pump, it
is possible that the hematocrit results will be reported
falsely high. If an instrument customized as “CPB, never
apply” is used for paents who are on the pump or have
decreased levels of protein, the hematocrit results may
be reported falsely low.
An instrument that is customized to prompt for CPB requires the
operator to answer Yes or No. Yes indicates that the CPB
correcon should be applied. No indicates that the CPB
correcon should not be applied. Select one of the following:
CPB, Prompt
Pr
ompts the operator to apply CPB correcon when the cartridge
has a hematocrit sensor.
CPB, always apply
Apply CPB c
orrecon every me a cartridge with a hematocrit
sensor is used.
CPB, never apply
Do not apply CPB c
orrecon when running a cartridge with a
hematocrit sensor.
186
i-STAT Alinity — System Operaons Manual Art: 746300-01 Rev. N Rev. Date: 02-Nov-2022

Table of Contents

Other manuals for Abbott i-STAT Alinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT Alinity and is the answer not in the manual?

Abbott i-STAT Alinity Specifications

General IconGeneral
BrandAbbott
Modeli-STAT Alinity
CategoryMedical Equipment
LanguageEnglish

Related product manuals